IM 12J05D02-01E
Parameter setting 5-19
Code Display
Function
Function detail
X
Y
Z
Default values
mA Output Functions
30
*mA
mA output
Output 1 is 0- 20 mA
0
1.1
Output 1 is 4- 20 mA
1
Output 2 is 0- 20 mA
0
Output 2 is 4-20 mA
1
31
*OUTP.F
Output function
Process value on mA1
0
0.2
Table on mA1
1
Process value on mA2
0
Table on mA2
1
Temperature on mA2
2
PI control on mA2
3
*D/R
Control action
Reverse acting PI-control
0
0
Direct acting PI control
1
32
*BURN
Burn-out
No burn-out function on mA1
0
0.0
Burn low on fail on mA1
1
Burn high on fail on mA1
2
No burn out function on mA2
0
Burn low on fail on mA2
1
Burn high on fail on mA2
2
33
*RGmA2
Prop. range
Proportional band: 0- 50 mg/l
5.0 mg/l
0-300 % saturation
34
*tImA2
Integral time
Integral time in seconds 10 - 1999 s
100 s
35
*TABL1
Table for mA1
21 point table: Process value at
0, 5, .. 100 % of span
none
36
*TABL2
Table for mA2
21 point table: 0, 5, .. 100 %
none
37
*DAMP
Output damping
damping 0- 120 s
0 s
38
*SWTCH
Range 1B
Disabled
0
0
Enabled
1
39
Not used
Note
To enable wash function, set Service Code 38 to 0, Service Code 42 to 4.0.0, and Service Code
51 to 0.1.0.