IM 12J05D02-01E

11-2 Appendix 2

2/4

QIS 12J05D02-01E

f. Press the [] key seven times. The data display will show “87” with the second digit

of 7 flashing.

g. Press the [ENT] key. The message display will show “PASS.”

(2) Setting Password 070

a. Press the [>] key once. The data display will show “000” with the second digit of 0

flashing.

b. Press the [] key seven times. The data display will show “070” with the second

digit of 7 flashing.

c. Press the [ENT] key. The data display will show “HIF.”

d. Press the [YES] key. The message display will show a manufacturing control

number.


This is the inspection mode.
Press the [ENT] key until the message display shows “mA1+2.”
Each time the [ENT] key is pressed, the value on the data display increases in steps of

4.0, like “0.0,” “4.0,” “8.0,” up to 20.0. Check the current outputs 1 and 2. Corresponding

to the data display, the current output must be within the range shown in Table 1.
When the data display shows “12.0,” pressing the [ENT] key causes the message display

to show “RIPPLE” and the data display to remain unchanged. Press the [ENT] key again.

The message display will show “mA” and the data display “16.0.”

Table 1

Data Display

Current Output 1 (mA DC)

Current Output 2 (mA DC)

4.0

4 ±0.02

4 ±0.02

12.0

12 ±0.02

12 ±0.02

20.0

20 ±0.02

20 ±0.02

3.4 Contact Operation Test

Following Section 3.3, press the [ENT] key until the message display shows “JUMPER.”

Then press the [ENT] key once, the message display will show “WAIT” and LEDs S1,

S2, WASH/S3, and FAIL/S4 flash rapidly. Check that the message display shows and

LEDs turns on/off in the order as shown in Table 2 and that a message of “FAIL x” does

not appear.

Table 2

Message Display

S1

S2

WASH/S3

FAIL/S4

LED1 OFF ON ON OFF
LED2 ON OFF ON OFF
LED3 ON ON OFF OFF
LED4 ON ON ON ON

3.5 Temperature

Indication Check

Following Section 3.4, press the [ENT] key until the message display shows “PT1000.” In

this state, change the resistance of the decade resistance box as shown in Table 3 and

check the data display. The value on the data display must be within the range shown in

Table 3.