IM 12J05D02-01E

Troubleshooting 8-3

Error Generation

Error content and Causes

Countermeasures

No. Mode

E10

All modes

EEPROM writing failure

Turn off the power and then turn it

Electronic circuit failure

back on and check whether or not the

system returns to normal.

If the failure occurs again, contact

Yokogawa to request repair.

E12

All modes

Abnormal measured value

Examine the setting for CODE 01, 02

The dissolved-oxygen value (%sat.) exceeds

10, 12 & 54.

150 % (or value set in CODE 54)

Accurately perform calibration.

E15

CODE 12

Temperature error correction failure

Accurately measure the temperature of

The difference between the corrected value

the measuring solution (dissolved oxygen

and the standard value exceeds the range

sensor).

corresponding to ±7.5°C (±13 5°F).

Examine whether the sensor cable fails

• Temperature entry is not accurate.

(partial disconnection or degraded

• Sensor cable failure

insulation) and then correct the

temperature error again.

E16

All modes

Call for maintenance-interval time exceeded

Perform maintenance. Reset interval.

E17

RANGE mode

Incorrect span.

Change the maximum or minimum value

In the range setting, a value which makes

in the range so that a span more than

the span less than 1.0 mg/l (or ppm) or 10

the permissible value is obtained.

% sat. has been set.

E18

CODE 35 & 36

Table values make no sense.

Reprogram.

E19

At data entry

Configuration failure

Re-enter a value within the permissible

A value not within the permissible range

range.

is entered.

E20

All modes

Instrument initialization failure

Repair is necessary. Contact Yokogawa

to request repair.

E21

All modes

EPROM comparison failure

Repair if necessary. Contact Yokogawa

to request repair.

E22

All modes

The time-out interval for high or low

Take measures depending on the purpose

alarm has elapsed.

to use this function.

(if the function is enabled in CODE 47)

Reset the error with the (

NO) key or

(

YES) key.

NOTE:

It is possible that the error message does not disappear immediately after corrective measures.

Then turn power off and the turn it back on and check whether the system returns to normal.