Yokogawa DO402 Dissolved Oxygen Analyzer User Manual
Page 81

IM 12J05D02-01E
Troubleshooting 8-3
Error Generation
Error content and Causes
Countermeasures
No. Mode
E10
All modes
EEPROM writing failure
Turn off the power and then turn it
Electronic circuit failure
back on and check whether or not the
system returns to normal.
If the failure occurs again, contact
Yokogawa to request repair.
E12
All modes
Abnormal measured value
Examine the setting for CODE 01, 02
The dissolved-oxygen value (%sat.) exceeds
10, 12 & 54.
150 % (or value set in CODE 54)
Accurately perform calibration.
E15
CODE 12
Temperature error correction failure
Accurately measure the temperature of
The difference between the corrected value
the measuring solution (dissolved oxygen
and the standard value exceeds the range
sensor).
corresponding to ±7.5°C (±13 5°F).
Examine whether the sensor cable fails
• Temperature entry is not accurate.
(partial disconnection or degraded
• Sensor cable failure
insulation) and then correct the
temperature error again.
E16
All modes
Call for maintenance-interval time exceeded
Perform maintenance. Reset interval.
E17
RANGE mode
Incorrect span.
Change the maximum or minimum value
In the range setting, a value which makes
in the range so that a span more than
the span less than 1.0 mg/l (or ppm) or 10
the permissible value is obtained.
% sat. has been set.
E18
CODE 35 & 36
Table values make no sense.
Reprogram.
E19
At data entry
Configuration failure
Re-enter a value within the permissible
A value not within the permissible range
range.
is entered.
E20
All modes
Instrument initialization failure
Repair is necessary. Contact Yokogawa
to request repair.
E21
All modes
EPROM comparison failure
Repair if necessary. Contact Yokogawa
to request repair.
E22
All modes
The time-out interval for high or low
Take measures depending on the purpose
alarm has elapsed.
to use this function.
(if the function is enabled in CODE 47)
Reset the error with the (
NO) key or
(
YES) key.
NOTE:
It is possible that the error message does not disappear immediately after corrective measures.
Then turn power off and the turn it back on and check whether the system returns to normal.