Yokogawa DO402 Dissolved Oxygen Analyzer User Manual

Page 63

background image

IM 12J05D02-01E

Parameter setting 5-27

Code Display

Function

Function detail

X

Y

Z

Default values

User Interface

50

*RET

Auto return

No return to meas. from HOLD

0

1

Return to meas after 10 min

1

51

*MODE

add. to MAINT

Setpoint adj. disabled

0

0.0.0

Setpoint adj. also in MAINT

1

Wash start disabled

0

Wash start in maintenance mode

1

Manual pressure disabled

0

Manual pressure enabled

1

52

*PASS

Password

Direct access to MAINT

0

0.0.0

MAINT protected by password

#

Direct access to COMM

0

COMM protected by password

#

Direct access to SERV

0

SERV protected by password

#

53

*Err.01

Error setting

Response & Time failure Soft fail

0

Response & Time failure Hard fail

1

1

Hard

*Err.02

Zero calibration

Soft fail

0

Zero calibration

Hard fail

1

1

Hard

*Err.03

Slope calibration

Soft fail

0

Slope calibration

Hard fail

1

1

Hard

*Err.04

Membrane failure

Soft fail

0

Membrane failure

Hard fail

1

1

Hard

*Err.07

Temperature sensor open Soft fail

0

Temperature sensor open Hard fail

1

1

Hard

*Err.08

Temp. sensor shorted

Soft fail

0

Temp. sensor shorted

Hard fail

1

1

Hard

*Err.09

Cell current abnormal

Soft fail

0

Cell current abnormal

Hard fail

1

1

Hard

*Err.12

Abnormal measured value Soft fail

0

Abnormal measured value Hard fail

1

1

Hard

*Err.16

Cell for maintenance

Soft fail

0

0

Soft

interval time exceeded

Cell for maintenance

Hard fail

1

interval time exceeded

*Err.22

Control time-out

Soft fail

0

0

Soft

Control time-out

Hard fail

1

*SOFT

LCD Fail Constant

0

0 LCD Fail

LCD only

1

54

*E12

overrange limit

E12 active at 0 - 300 % sat (Note)

150%

55

*CALL.M

E16 setting

No call for maintenance

0

0

Call for maint. enabled

1

E16 message after 1- 250 days

30

56

*UNIT

PV unit

Unit of measurement mg/l

0

0

Unit of measurement ppm

1

Unit of measurement % sat.

2

57-59

Not used

Pass: 111 for #=1,

333 for #=2,

777 for #=3,

888 for #=4,

123 for #=5,

957 for #=6,

331 for #=7,

546 for #=8,

847 for #=9

Note: For setting a value greater than 199.9, move the decimal point. For a value greater than 200%,

decimal places cannot be used.