Yokogawa DO402 Dissolved Oxygen Analyzer User Manual
Page 51

IM 12J05D02-01E
Parameter setting 5-15
Code
Display
Function
Function detail
X
Y
Z
Default values
Parameter Specific functions
01
*S. TYPE Sensor type
Galvanic
0
0
Polarographic
1
*I.CELL
Sensor output
0
0
Galvanic
1
User defined
9
3.75 μA
*I.CELL
Sensor output
50 nA
0
0
Polarographic
User defined
9
50 nA
02
*CHECK
Sensor check
Zero check: disabled
0
0.1.0
Zero check: enabled
1
Slope check: disabled
0
Slope check: enabled
1
Membrane check: disabled/only for Galvanic
0
Membrane check: enabled/only for Galvanic
0 to 19.99 μA for Galvanic
*ZERO
Maximum permissible
0 to 199.9 nA for Polarographic
Zero offset
0 to 99.9%
*SL.MIN
100 to 400%
0.5 μA
5 nA
40%
200%
*SL.MAX
NOT USED
Minimum calibrated slope
Maximum calibrated slope
1
(*1)
03
NOT USED
04
*SAL.TY
Salinity comp.(*2)
Compensation disabled
0
0
Compensation enabled
1
*[CL-]
NaCl conc.
Concentration in g/kg water:
range 0.1 to 99.9 g/kg
Valid when code
04 is 1
05-09
Not used
3.75 μA for 50 micron type
7.50 μA for 25 micron type
T0502.EPS
(*1): Membrane check can be enabled when galvanic sensor is selected and cannot be used unless the
floating ball holder, PB350G or PB360G, is used.
(*2): The instrument will not perform salinity compensation simply by setting the salinity compensation to
“1” (enabled) in Code 04, but requires the NaCl concentration. For measurement in consideration of
the salinity of a sample, first, manually calibrate with values corrected for salinity and enter the NaCl
concentration in Code 04. (See Section 6-4, “Calibration method using manual calibration method”)
(*3): Set 60 nA for the DO70G optical dissolved oxygen sensor.