beautypg.com

5 cleaning and replacing the capsule assembly, Cleaning and replacing the capsule assembly -5, Important – Yokogawa EJX115A User Manual

Page 42: Caution

background image

<8. Maintenance>

8-5

IM 01C25K01-01E

IMPORTANT

Exercise care as follows when cleaning the

manifold assembly.

• Handle the manifold assembly with care, and

be careful not to damage the inner part of the

manifold, spacer, and orifice. Be especially

careful not to damage or distort the orifice

edge (orifice bore).

• Do not use a chlorinated or acidic solution for

cleaning.

• Rinse thoroughly with clean water after

cleaning.

■ Reassembling the Manifold Assembly
1) Reassemble the orifice gasket, orifice, and

spacer into the manifold in that order.

When reassembling, refer to Figure 8.5 to

ensure that they are placed in the correct

direction.

Replace the orifice gasket with a new gasket.

2) Mount the process connector as shown in

Subsection 9.4.3.

3) Mount the manifold on the cover flange with the

four bolts. Tighten the four bolts uniformly to a

torque of 39 to 49 N·m {4 to 5 kgf·m}.

Replace the manifold gaskets with new

gaskets.

4) After completing reassembly, a leak test

must be performed to verify that there are no

pressure leaks.

NOTE

Exercise care as follows when reassembling the

manifold assembly. (See Figure 8.5.)

• Be careful not to reassemble the orifice in

the wrong direction. Note that the spacer is

configured so that it cannot be placed in the

reverse direction.

• When mounting the manifold on the cover

flange, confirm the indication “flow direction”

shown on the manifold surface and the high

and low pressure sides of the pressure-

detector section.

Mount the manifold so that the upstream

side of process fluid flow is located at the

high pressure side of the pressure-detector

section.

F0805.ai

Process connector

gasket

Process connector

gasket

Spacer

Orifice

Bolt

Orifice gasket

Manifold gasket

Cover flange

Manifold

Low pressure side

High pressure side

Figure 8.5

Manifold Assembly

8.4.5 Cleaning and Replacing the Capsule

Assembly

This subsection describes the procedures for

cleaning and replacing the capsule assembly. (See

figure 8.6.)

CAUTION

Cautions for Flameproof Type Transmitters

Users are prohibited by law from modifying the

construction of a flameproof type transmitter. If

you wish to replace the capsule assembly with

one of a different measurement range, contact

Yokogawa.

The user is permitted, however, to replace a

capsule assembly with another of the same

measurement range. When doing so, be sure to

observe the following.

• The replacement capsule assembly must

have the same part number as the one being

replaced.

• The section connecting the transmitter and

capsule assembly is a critical element in

preservation of flameproof performance, and

must be checked to verify that it is free of

dents, scratches, and other defects.

• After completing maintenance, be sure to

securely tighten the setscrews that fasten

the transmitter section and pressure-detector

section together.