beautypg.com

A2-5 – Yokogawa EJA440E User Manual

Page 95

background image

A2-5

IM 01C25T01-06EN

Checklist (5/5)

No.

Items

Parameters

Result

Example

10-b

Simulation of Blockage detection operation

• Close completely the valve for the side where

the alarm of Invalid Reference Value is not

generated.

For the case that the high-pressure side value is

closed;

• Record the values of

fDP, fSPl, fSPh, BlkF,

and

DPAvg after the certain time,

(

Diag Period X Diag Supp Count), passed.

*: Record the value after checked that the

status is “GOOD”.

fDP*

7.48562E-09

fSPh*

7.14085E-09

For the case that the low-pressure side value is

closed;

• Record the values of

fDP, fSPl, fSPh, BlkF,

and

DPAvg after the certain time,

(

Diag Period X Diag Supp Count), passed.

*: Record the value after checked that the

status is “GOOD”.

fDP*

7.48562E-09

fSPI*

7.23277E-09

• Record the status of Checkbox in

Diag

Option.

• Check that the alarms status of “B Blocking” is

set.

Note:

If the alarm of “ILBD over range” is generated,

the valve may be closed too much tightly. Open

valve a little and record the updated status of the

parameters.

Diag Option

A Blocking

Large Fluct L

Large Fluct H

L Side Blocking

H Side Blocking

B Blocking

Invalid Ref F

Invalid Ref SPH

Invalid Ref SPL

Invalid Ref DP

ILBD over range

• Check that the alarm of “B Blocking” is not

generated.

Diag Error

B Blocking

This manual is related to the following products: