12 ilbd parameter list, 12 ilbd parameter list -19 – Yokogawa EJA440E User Manual
Page 66
<4. Diagnostics>
4-19
IM 01C25T01-06EN
4.2.2.12 ILBD Parameter List
# Parameter name Default value
Explanation
1 Diag Error
0x0000
The results detected by ILBD or Heat trace monitoring are stored into this
parameter.
Also the condition abnormality in the diagnostic process is stored as an error.
2 Diag Option
0x08FC
The masking in this parameter enable to display each error message and the
status to the output signal or LCD.
The error assigned to each bit is corresponding to that of
Diag Error.
Writable only when Diag Mode is "Stop".
3 Diag Out Option Off
Output mode of 4-20mA when an advanced diagnostic alarm is generated.
There are following three output modes; Off, Burnout, or Fall back.
4 Diag Fixed Out
Val
21.6 mA
Parameter for "Fall back" function in the Diag Out option. The output value of
4-20 mA analog signal is specified when an alarm is generated. The value can be
entered within 3.6 to 21.6 mA.
5 DO Select
Off
The variables for status output are specified to this parameter. When the
advanced diagnostic function (option code /DG6) is installed, the parameters
monitoring in diagnostic process can be also assigned to the status output.
6 Diag Mode
Stop
The operation mode of ILBD is set.
Stop:
The blockage detection is stopped.
Calculate: The blockage detection is carried out.
The alarms are generated along with the detected result.
Reference: The reference values are obtained and the update values are
overwritten.
After setting, this mode moves automatically to "Calculation".
7 Diag Period
180 (s)
The data acquisition period for ILBD is set within 20 to 65535 (s). If the process
fluctuation values are unsteady, this value is changed to the longer to enhance
the accuracy of the blockage detection.
Writable only when Diag Mode is "Stop".
8 Diag Supp Count 3
Detection count to generate an alarm. When the statistical value as
Ratio fDP
and
BlkF exceeds consecutively the threshold by number of times preset to this
parameter, it is estimated that the impulse line is plugged.
9 Diag Description
Memo field. 32 alphanumerics
10 fDP
Average value of the sum of squares of differential pressure fluctuation.
11 fDP Status
Status of
fDP
12 fSPl
Average value of the sum of squares of low-pressure-side static pressure
fluctuation. For gauge/absolute pressure transmitter, 0 is set.
13 fSPl Status
Status of
fSPl
14 fSPh
Average value of the sum of squares of high-pressure-side static pressure
fluctuation. For gauge/absolute pressure transmitter, 0 is set.
15 fSPh Status
Status of
fSPh
16 BlkF
Blockage degree characterized in comparison of high- and low-pressure side
pressure fluctuation value.
17 BlkF Status
Status of
BlkF
18 DPAvg
Ratio of the average of differential pressure/pressure to the maximum span of an
EJX transmitter.
19 DPAvg Status
Status of
DPAvg
20 Ratio fDP
CRatio fDP or NRatio fDP is used by Diag Comp setting.
fDP decreases and this parameter is used to determine whether one or both
sides are plugged.
21 Ratio fDP Status
Status of
Ratio fDP
22 Ratio fSPl
SQRT (
fSPl/Ref fSPl).
fSPl decreases and this parameter is used to determine whether low-pressure-
side is plugged.
23 Ratio fSPl Status
Status of
Ratio fSPl
24 Ratio fSPh
SQRT (
fSPh/Ref fSPh).
fSPh decreases and this parameter is used to determine whether high-pressure-
side is plugged.
25 Ratio fSPh
Status
Status of
Ratio fSPh
26 Ref fDP
Value of
fDP obtained under normal condition.
27 Ref fDP Status
Status of
fDP obtained under normal condition.