Sync 120j sel, 2 charge defibrillator – ZOLL R Series Monitor Defibrillator Rev K Operators Guide User Manual
Page 66
![background image](/manuals/355925/66/background.png)
C
HAPTER
5
S
YNCHRONIZED
C
ARDIOVERSION
5–4
www.zoll.com
9650-0912-01 Rev. K
A SYNC XXXJ SEL. message appears on the display. If DEFIB XXXJ SEL. appears, press the
Sync On/Off softkey. (If your unit supports Remote Sync, you must also press the Sync
softkey.) Two quick beeps sound.
Unless otherwise configured, the unit automatically exits Sync mode after each shock and when
the Mode Selector is moved to MONITOR, PACER or OFF.
To reactivate Sync mode, press the Sync On/Off softkey again. (If your unit supports Remote
Sync, press the Sync softkey again.) Changing the selected energy levels does not cause the
unit to leave Sync mode.
Note that the unit can be configured to stay in Sync mode after defibrillation, if desired. Refer
to the R Series Configuration Guide for instructions.
2 Charge Defibrillator
Press the CHARGE button on the front panel or on the apex paddle handle.
To abort charging and increase or decrease the selected energy after the CHARGE button has
been pressed, use the ENERGY SELECT buttons on either the defibrillator front panel or the
sternum paddle. Press the CHARGE button again to charge the unit to the newly selected
energy level.
After charging the unit to the selected energy, either the front panel SHOCK button or the
APEX paddle charge indicator illuminates. A distinctive audible tone sounds and the SYNC
XXXJ READY message is displayed.
The defibrillator is now ready to deliver therapy.
SYNC 120J SEL.
Code
Marker
Options
00:01
ECG
Param
Report
Data
Sync
On/Off
Alarms
3
SHOCK
SHOCK
ENERGY
SELECT
ENERGY
SELECT
ANALYZE
ANALYZE
CHARGE
CHARGE
2
1
RELEASE
RELEASE
or