Recorder
9650-0912-01 Rev. K
ZOLL R Series Operator’s Guide
13–3
Recorder
6. Noisy ECG, artifact, wandering
baseline
•
Consider 1 – 21Hz filter bandwidth (see R Series
Configuration Manual).
•
Prepare the patient’s skin prior to electrode attachment.
•
Check for proper adhesion of electrodes to patient.
•
Reduce or eliminate ECG artifact due to electrode or
patient cable movement. Route cables so that they don’t
pull on electrodes or swing excessively.
•
Ensure patient is motionless.
•
Check for possible excessive radio frequency
interference.
7. Poor ECG signal amplitude,
calibration pulse normal
•
Select another lead.
•
Apply new electrodes using different placement.
8. Inconsistent QRS beep or heart
rate
•
Select another lead.
•
Alter ECG electrode placement.
9. Sync marker is absent or
inconsistent with QRS waveform
on display and recorder
•
Ensure device is in Sync mode.
•
Ensure that device in NOT in Remote Sync mode.
•
Change ECG lead selection.
•
Alter ECG electrode placement.
•
Paper too narrow. It should be 90 mm wide.
10. No ECG trace or dashed line on
display
•
Device is in Remote Sync mode. Press the
Sync On/Off softkey to exit Remote Sync mode.
Symptom
Recommended Action
1. CHECK RECORDER message
•
Recorder out of paper.
•
Remove paper, check paper type, check recorder for
paper jam, and reload paper.
•
Recorder door is open.
•
Paper is loaded upside down.
2. Recorder makes stuttering sound
when activated.
•
Check recorder for paper jam.
3. Light or poor quality printing
•
Ensure correct paper is in use.
•
Ensure paper is installed grid side against recorder print
head.
•
Recorder print head requires cleaning.
4. Summary Report will not print
when Report/Print Chart softkeys
are activated.
•
15 seconds have not elapsed since one of the events
that trigger Summary Report to record have occurred.
Wait 15 seconds, then try again.
Symptom
Recommended Action