N = 6: Missing required configuration data for Network
Profile 1 with Enterprise Authentication selected
(Data COMM II only)
Action: Check the following:
If Authentication Protocol is set to PEAP, make sure
that the following are configured:
•
User Identity
•
User Password
If Authentication Protocol is set to TLS, make sure
that the following are configured:
•
User Identity
•
Private Key Password
•
Client Certificate
N = 7: Missing required configuration data for Network
Profile 2 with Enterprise Authentication selected
(Data COMM II only)
Action: Same as above.
N = 8: The Defib Upload Server's Username /
Password are blank within the Full Disclosure and
Defib History server configuration items
Data COMM II only)
Action:
If Full Disclosure Transfer Mode is set to Server, make
sure that the Username (Full Disclosure) and
Password (Full Disclosure) entries are valid.
If Defib History Transfer Mode is set to Manual,
Automatic, or Both, make sure that the Username
(Defib History) and Password (Defib History) entries
are valid.
N = 9: Missing required certificate file for Network
Profile 1 with Enterprise Authentication selected
(Data COMM II only)
Action:
If Authentication Protocol is set to PEAP, make sure
that a root certificate has been added (if not using the
default ZOLL root certificate).
If Authentication Protocol set to TLS, make sure that a
root certificate has been added (if not using the
default ZOLL root certificate) and that the correct
client certificate has been configured.
N = 10: Missing required certificate file for Network
Profile 2 with Enterprise Authentication selected
(Data COMM II only)
Action: Same as above.