Full disclosure recording, Incident logs, Printing a partial summary report – ZOLL R Series Monitor Defibrillator Rev K Operators Guide User Manual
Page 108: Printing an incident log

C
HAPTER
10
E
VENT
R
ECORDS
AND
R
EPORTS
10–8
www.zoll.com
9650-0912-01 Rev. K
Printing a Partial Summary Report
To print a partial summary report:
1. Press the Report Data softkey, then press Print Chart.
2.
Press the Print Range softkey.
3.
Use the First Event, Prev. Event, Next Event and Last Event softkeys to locate the event
from which printing will start (all subsequent events are also printed).
4.
Press the Print softkey.
The unit prints all records from the selected item to the most recent.
Note:
The overview information (such as the number of shocks delivered) covers all stored
summary data, not just the selected range.
Full Disclosure Recording
Along with event information captured in Summary Report, R Series also records the full
disclosure CPR sensor and parameter waveforms. Full disclosure recordings are erased at the
same times as Summary Reports.
Incident Logs
An incident log is an abbreviated list of all major events recorded in summary report. You can
print an incident log that includes the following events and their time of occurrence.
•
Unit powered on.
•
Defibrillation advisory message issued (for example, CHECK PATIENT or SHOCK
ADVISED)
•
Shock delivered (and energy level)
•
PACER mode selected
•
Alarm triggered
•
Stripchart printing started
•
Code marker entered
In addition, the incident log lists the following:
•
System serial number
•
Device identification number
•
Report start time (when the summary data was last erased)
•
Time of the last recorded event
•
Total number of shocks delivered
•
Total cumulative pacing time
Printing an Incident Log
To print an incident log, press the Report Data softkey, then press the Print Log softkey.
The log is printed on the stripchart, starting with the oldest entry.