R Series Rectilinear Biphasic Waveform Characteristics
9650-0912-01 Rev. K
ZOLL R Series Operator’s Guide
A–13
R Series Rectilinear Biphasic Waveform
Characteristics
Table A-1 shows the characteristics of the R Series Rectilinear Biphasic™ waveform when
discharged into 25 ohm, 50 ohm, 100 ohm, 125 ohm, 150 ohm and 175 ohm loads at the
maximum energy setting of 200 joules.
Table A-1.
R Series Rectilinear Biphasic Waveform Characteristics
200 J discharged into
25
50
100
125
150
175
First phase
Maximum initial current
31.4 A
30.4 A
19.7 A
19.4 A
16.7 A
15.6 A
Average current
27.1 A
24.9 A
17.5 A
16.2 A
14.4 A
13.2 A
Duration
6 ms
6 ms
6 ms
6 ms
6 ms
6 ms
Interphase duration
(between first and
second phases)
200
s
200
s
200
s
200
s
200
s
200
s
Second phase
Initial current
29.2 A
18.8 A
15.1 A
13.2 A
12.1 A
11 A
Average current
14.7 A
13 A
12.5 A
11.3 A
10.7 A
9.9 A
Duration
4 ms
4 ms
4 ms
4 ms
4 ms
4 ms
Table A-2.
Delivered Energy at Every Defibrillator Setting into a Range of Loads
Selected
Energy
Load
25
50
75
100
125
150
175
Accuracy*
1 J
1 J
1 J
1 J
1 J
1 J
1 J
1 J
±15%
2 J
1 J
2 J
2 J
2 J
2 J
2 J
2 J
±15%
3 J
2 J
3 J
3 J
3 J
3 J
3 J
3 J
±15%
4 J
3J
4 J
4 J
5 J
5 J
5 J
4 J
±15%
5 J
3 J
5 J
6 J
6 J
6 J
6 J
6 J
±15%
6 J
4 J
6 J
7 J
7 J
7 J
7 J
7 J
±15%
7 J
5 J
7 J
8 J
8 J
8 J
8 J
8 J
±15%
8 J
5 J
8 J
9 J
9 J
10 J
9 J
9 J
±15%
9 J
6 J
9 J
10 J
11 J
11 J
11 J
10 J
±15%
10 J
7 J
10 J
12 J
12 J
12 J
12 J
12 J
±15%
15 J
10 J
16 J
17 J
18 J
18 J
18 J
17 J
±15%
20 J
14 J
21 J
23 J
24 J
24 J
24 J
23 J
±15%
30 J
21 J
32 J
35 J
36 J
37 J
36 J
35 J
±15%
50 J
35 J
54 J
59 J
61 J
62 J
61 J
59 J
±15%
70 J
49 J
76 J
83 J
85 J
87 J
86 J
83 J
±15%
75 J
53 J
81 J
89 J
91 J
93 J
92 J
89 J
±15%
85 J
60 J
92 J
101 J
104 J
106 J
104 J
101 J
±15%
100 J
71 J
109 J
119 J
122 J
125 J
123 J
119 J
±15%
120 J
85 J
131 J
143 J
147 J
150 J
147 J
143 J
±15%