Chapter 11 file transfer, Transferring files to an external device – ZOLL R Series Monitor Defibrillator Rev K Operators Guide User Manual
Page 111

9650-0912-01 Rev. K
ZOLL R Series Operator’s Guide
11–1
Chapter 11
File Transfer
This chapter describes procedures for transferring files from the R Series defibrillator to an
external system, such as a personal computer or handheld device. It also explains how to
remove, install and erase a Compact Flash card.
Transferring Files to an External Device
You can transfer the following files from the R Series defibrillator to an external device:
•
Device check, activity log, and full disclosure waveforms
•
Defibrillator test information
The unit includes the following data transfer features:
•
802.11 Wi-Fi (optional)
•
USB device connector (optional)
•
Compact flash card slot
To retrieve and review event files, you need one of the following software packages installed on
the receiving equipment:
•
ZOLL CodeNet Central software
•
ZOLL RescueNet Code Review software
To retrieve and review maintenance files, you need ZOLL Defib Dashboard software installed
on the receiving equipment.
R Series defibrillators use the Microsoft Windows file structure for stored records. Files can be
transferred to a properly equipped Windows-based personal computer or handheld device. With
CodeNet Central software, the personal computer allows the user to access the files for review.