Default Cuff Inflation Pressure:
Adult: 160 mmHg
Pediatric:120 mmHg
Neonatal:90 mmHg
Maximum Cuff Inflation Pressure:
Adult: 270 mmHg
Pediatric:170 mmHg
Neonatal:130 mmHg
Single Fault Backup Overpressure Limit:
Adult: 308 mmHg
Pediatric:205 mmHg
Neonatal:154 mmHg
Typical Determination time without Artifact:
Measurements on the deflation: 30 to 45 seconds
Measurements on the inflation (SureBP)*: 15 to 30 seconds
* using dual lumen cuffs
Maximum Determination Time - Measurement on the Inflation
Adult: 150 seconds
Pediatric:120 seconds
Neonatal:80 seconds
Blood Pressure Validation:
Blood pressure measurements determined with this device are equivalent to those obtained by a
trained observer using the cuff/stethoscope auscultation method for adults and pediatric patients
and equivalent to an intra-arterial measurement for neonatal patients, within the limits
prescribed by the American National Standards Institute (ANSI-AAMI SP10). To receive a
copy of the report containing the AAMI SP10 results, contact the ZOLL Technical Service
Department.
Invasive Pressures
Number of Channels: 3
Pressure range: -30 to 300 mmHg
Pressure Accuracy: +/- 2 mmHg or 2% of reading, whichever is greater, plus transducer error.
Pulse Rate Range: 25 to 250 BPM
Pulse Rate Accuracy: +/- 3 BPM, or +/- 3% of value whichever is greater
Pulse Rate Display: Average of last 4 beat-to-beat intervals.
Zero Adjust: + / - 200 mmHg