Battery, General – ZOLL Propaq M Rev B User Manual
Page 156

Specifications
A-4
www.zoll.com
9650-0803-01 Rev. B
Physiological Alarms (NIBP, SpO2, Resp, CO2, IP & Temp):
Audible: Same as Heart Rate Alarm
Visual: Physiological alarms cause the alarming parameter to be displayed in Red with a white
background. The red device status LED will flash at a rate of 1.7 Hz.
Silence Duration: 90 seconds.
Invalid Operation Alert Tone:
A short, low-pitched tone is audible when a selected control button is unavailable for use or an
invalid entry is detected. Tone frequency is 160 Hz. Duration is 250 msec.
Delays (hold off time):
Heart rate/pulse rate:
•
i
f source is ECG, 5 seconds
•
if source is SpO2, 10 seconds
•
if source is IBP, 5 seconds
•
if source is NIBP, no hold off
SpO2 Saturation: 10 seconds
EtCO2: 5 seconds
FiCO2: 5 seconds
IBP (Systolic, Diastolic, Mean): 2 seconds
Temperature: 2 seconds
Battery
Type: Rechargeable Lithium-Ion, 11.1Vdc, 6.6 Ah, 73Wh
Capacity:
With a new, fully charged battery operating at room temperature:
•
At least
7.5 hours of continuous monitoring of ECG, SpO2, CO2, three Invasive Pressure
channels, and 2 channels of Temperature, with NIBP measurements every 15 minutes
(display set to 30%).
Note:
Proper battery care is required to maintain maximum available capacity.
Battery Indicators:
5 Battery capacity LED indicators, Fault indicator, Recalibration indicator
Recharge Rate: 100% in 4 hours, when initiated at Low Battery indication.
General
Weight:
Without Battery: 7.2 lbs.
With Battery: 8.5 lbs.
Dimensions:
Without Handle: 8.9” x 8.7” x 6.5”
With Handle: 8.9” x 10.4” x 7.0”