Pulse oximeter, Non-invasive blood pressure – ZOLL Propaq M Rev B User Manual
Page 159

Pulse Oximeter
9650-0803-01 Rev. B
Propaq M Operator’s Guide
A-7
Pulse Oximeter
Measurement Range: 0 to 100%
Accuracy: 70 to 100% +/- 3%, 0 to 70% unspecified
Note:
Since pulse oximeter equipment measurements are statistically distributed, only two-
thirds of pulse oximeter equipment measurements can be expected to fall within +/-3%
of the value measured by a CO-oximeter.
Saturation Update Period: Updated once per second
SpO2 Saturation Alarm Hold Off: 10 seconds
Pulse Rate Range: 20 to 250 BPM ± 3% or 3 BPM whichever is greater
Pulse Rate Update Period: Updated once per second
Pulse Rate and Saturation Display:
Normal Response: Average of pulse rate and saturation signal over 6 to 7 seconds.
Wavelengths: Approximately 660nm and 900nm
Displayed Data:
Numeric blood oxygen saturation value
Numeric pulse rate
Pulse Amplitude bar indicator
Plethysmographic waveform
Probe Compatibility: DS-100A, D-25, R-15, D-25L, D-20, EC-4, EC-8, RS-10
Power Output: < 15 mW
Non-Invasive Blood Pressure
Technique: Non-invasive oscillometric method
Operating Modes: Automatic and manual
Automatic Intervals: 1, 2, 3, 5, 10, 15, 30 and 60 minute intervals.
Turbocuf: Maximum measurements allowable in a 5 minute period
Pressure Measurement Range:
Systolic: 20 to 260 mmHg
Diastolic:10 to 220 mmHg
Mean: 13 to 230 mmHg
Static Pressure Accuracy: +/- 3 mmHg
Pulse Rate Range:
Adult:
30 to 200 +/- 5 BPM
Pediatric:30 to 200 +/- 5 BPM
Neonatal:35 to 220 +/- 5 BPM