Pulse Oximeter

9650-0803-01 Rev. B

Propaq M Operator’s Guide

A-7

Pulse Oximeter

Measurement Range: 0 to 100%

Accuracy: 70 to 100% +/- 3%, 0 to 70% unspecified

Note:

Since pulse oximeter equipment measurements are statistically distributed, only two-
thirds of pulse oximeter equipment measurements can be expected to fall within +/-3%
of the value measured by a CO-oximeter.

Saturation Update Period: Updated once per second

SpO2 Saturation Alarm Hold Off: 10 seconds

Pulse Rate Range: 20 to 250 BPM ± 3% or 3 BPM whichever is greater

Pulse Rate Update Period: Updated once per second

Pulse Rate and Saturation Display:

Normal Response: Average of pulse rate and saturation signal over 6 to 7 seconds.

Wavelengths: Approximately 660nm and 900nm

Displayed Data:

Numeric blood oxygen saturation value

Numeric pulse rate

Pulse Amplitude bar indicator

Plethysmographic waveform

Probe Compatibility: DS-100A, D-25, R-15, D-25L, D-20, EC-4, EC-8, RS-10

Power Output: < 15 mW

Non-Invasive Blood Pressure

Technique: Non-invasive oscillometric method

Operating Modes: Automatic and manual

Automatic Intervals: 1, 2, 3, 5, 10, 15, 30 and 60 minute intervals.

Turbocuf: Maximum measurements allowable in a 5 minute period

Pressure Measurement Range:

Systolic: 20 to 260 mmHg

Diastolic:10 to 220 mmHg

Mean: 13 to 230 mmHg

Static Pressure Accuracy: +/- 3 mmHg

Pulse Rate Range:

Adult:

30 to 200 +/- 5 BPM

Pediatric:30 to 200 +/- 5 BPM

Neonatal:35 to 220 +/- 5 BPM