Setting co, And respiration rate alarms, Enabling/disabling alarms and setting co – ZOLL Propaq M Rev B User Manual
Page 114: Alarm limits

C
HAPTER
10 M
ONITORING
CO2
10-8
www.zoll.com
9650-0803-01 Rev. B
Setting CO
2
and Respiration Rate Alarms
The Propaq M unit sounds alarms whenever measurements are outside set limits for the
following:
•
High and Low EtCO
2
•
High and Low Respiration Rate (in Breaths/Minute)
•
High FiCO
2
•
No Breath Alarm
Enabling/Disabling Alarms and Setting CO
2
Alarm Limits
You can enable (or disable) CO
2
alarms and set upper and lower alarm limits through the
Alarms quick access key (
) or the CO2 Parameter Control Panel.
To configure CO
2
alarms through the Alarms quick access key:
1. Press the More quick access key (
).
2.
Press .
3.
Press the Limits quick access key (
).
4.
Use the navigation buttons to select appropriate alarm menu. For CO
2
, the alarm menus are:
EtCO2 Alarm, FiCO2 Alarm, or RR/BR Alarm.On the alarm settings menu, use the
navigation keys to select the fields that you want to change. The fields are
•
Upper Limit Enable
•
Lower Limit Enable (only EtCO2 and RR/BR Alarms)
•
Upper Limit
•
Lower Limit (only EtCO2 and RR/BR Alarms)
5.
When you are finished changing values on the alarm menu, select the Backarrow key to exit
the menu.
CO
2
CO
2
Limits