beautypg.com

ZOLL Propaq M Rev A User Manual

Page 62

background image

C

HAPTER

4

L

IFE

C

YCLE

C

OUNTERS

58

9650-0804-01 Rev. A

CO2

On the Propaq M, do the following:

1. Press the Setup quick access key (

).

2.

Use the navigation keys to select Supervisor>Service>Diagnostic>CO2>Service Codes.

3.

Select Reset CRC Counter. Select Yes when prompted to confirm the reset.

See “Functional Verification Tests” on page 1-22 to perform the functional verification tests.

SpO2

On the Propaq M, do the following:

1. Press the Setup quick access key (

).

2.

Use the navigation keys to select Supervisor>Service>Diagnostic>SpO2>Error Counts.

3.

Select Reset CRC Counter. Select Yes when prompted to confirm the reset.

See “Functional Verification Tests” on page 1-22 to perform the functional verification tests.