Functional verification tests, Before you begin the functional verification tests – ZOLL Propaq M Rev A User Manual
Page 26
C
HAPTER
1
M
AINTENANCE
T
ESTS
22
9650-0804-01 Rev. A
Functional Verification Tests
Before You Begin the Functional Verification Tests
•
Assemble the tools listed in following section, “Equipment You Need to Perform the
Functional Verification Tests.”
•
Keep an extra fully charged Propaq M battery available.
•
Schedule an hour to conduct the entire maintenance test.
•
Perform the tests in the order presented.
•
Perform all the steps of each test procedure.
•
Complete all the steps of the procedure before evaluating the test results.
Equipment You Need to Perform the Functional Verification Tests
For testing purposes, you can substitute an equivalent device.
•
Fluke Impulse 7000DP Analyzer, or equivalent, with appropriate interface adapter
(Dynatech Nevada Inc. part number 3010-0378)
•
IBP/Temp Simulator, or equivalent
•
NIBP Simulator
•
Single or dual lumen NIBP hose
•
Adult NIBP cuff and a rigid cylinder to wrap it around
•
5% CO
2
calibration gas cylinder with CO
2
sample line (kit available from Scott Medical
Products, PN 4653ORF-BD)
•
ECG Simulator; 12 Lead Simulator for 12 Lead test (e.g., Symbio CS1201)
Accessories You Need for the Propaq M Options Functional Verification
Tests
•
3-lead, 5-lead and 12-lead ECG cables. (12-lead cable needed if 12-lead option is
installed.)
•
NIBP cuff and hose
•
IBP cable
•
SpO
2
cable
and sensor
•
EtCO
2
cable
and sensor
•
Paddles
•
Battery
•
Auxiliary power source