Propaq M Error Messages

9650-0804-01 Rev. A

ZOLL Propaq M Service Manual

35

TEMP PROBE2 SHORTED

An A/D value was detected
that indicated the probe was
shorted.

Replace the temperature
probe or adapter cable. If
problem persists, replace
the right side panel.

TEMP1 CIRCUIT FAILURE

The software detected an
out-of-range A/D converter.

Remove probe, power
cycle. If problem persists,
replace main board.

TEMP2 CIRCUIT FAILURE

The software detected an
out-of-range A/D converter.

Remove probe, power
cycle. If problem persists,
replace main board.

IBP DEVICE FAILURE

IBP software detected a
problem with the IBP data
that is most likely a hardware
issue.

Replace the main board.

IBP PROBE1 SHORTED

An A/D value was detected
that indicated the probe was
shorted.

Replace the IBP probe. If
problem persists, replace
the right side panel.

IBP PROBE2 SHORTED

An A/D value was detected
that indicated the probe was
shorted.

Replace the IBP probe. If
problem persists, replace
the right side panel.

IBP PROBE3 SHORTED

An A/D value was detected
that indicated the probe was
shorted.

Replace the IBP probe. If
problem persists, replace
the right side panel.

IBP PROBE1 CIRCUIT
FAILURE

An unsupported probe was
inserted or there is a circuit
problem.

Unplug probe, power cycle.
If problem persists, replace
main board.

IBP PROBE2 CIRCUIT
FAILURE

An unsupported probe was
inserted or there is a circuit
problem.

Unplug probe, power cycle.
If problem persists, replace
main board.

IBP PROBE3 CIRCUIT
FAILURE

An unsupported probe was
inserted or there is a circuit
problem.

Unplug probe, power cycle.
If problem persists, replace
main board.

SPO2 DEVICE FAILURE

Communication failure or
device malfunction.

Check SpO

2

communication cables. If
problem persists, replace
the SpO

2

module.

Error Message

Explanation

Technical Action