beautypg.com

Ibp test – ZOLL Propaq M Rev A User Manual

Page 33

background image

Functional Verification Tests

9650-0804-01 Rev. A

ZOLL Propaq M Service Manual

29

IBP Test

Tools Needed

IBP patient simulator (with accuracy better than +/- 1 mmHg), or equivalent

Setup

Connect the test cables to IBP patient simulator.

Do This...

Observe This...

Pass/Fail

1.

Connect the IBP patient simulator to the
Propaq P1 channel.

2.

Set the IBP patient simulator to 0 mmHg.

3.

On the Propaq M, press > IBP.

4.

On the Propaq M, Press > P1->0<-.

Display of 0/0(+/-2) in the P1 box
at the bottom of the display (P1 is
zeroed).

The waveform for channel 1 is
present.

o

o

o

o

5.

Set the IBP patient simulator to 200
mmHg (or as close to 200 mmHg as your
simulator allows).

Display of 200/200(+/-5) in the P1
box at the bottom of the display as
xxx/yyy where yyy=diastolic (P1
diastolic pressure is 200 (+/-5)).

o

o

6.

Repeat steps 1 through 5 of this procedure
for P2 and P3 channels.