beautypg.com

Test – ZOLL Propaq M Rev A User Manual

Page 31

background image

Functional Verification Tests

9650-0804-01 Rev. A

ZOLL Propaq M Service Manual

27

SpO

2

Test

Tools Needed

SpO

2

Simulator: Fluke Index 2 or equivalent, Nellcor

TM

SpO

2

finger probe

Test Setup

Disconnect any ECG cable or IBP probes from the Propaq M.

Do This...

Observe This...

Pass/Fail

1.

Power on the Propaq M without the

SpO

2

sensor attached.

Verify that the

SpO

2

numeric

displays INITIALIZING, changes
to SEARCHING, then goes blank.

2.

Connect the

SpO

2

sensor to the Propaq M.

Verify that

SpO

2

numeric

displays SEARCHING.

3.

Set the simulator to 70%

SpO

2

and 60 PPM.

Attach the

SpO

2

sensor to the simulator.

Verify that after a brief period of
initialization, the

SpO

2

numeric

displays an

SpO

2

value between

67% and 73%.

Verify that the displayed pulse
rate reads 60 +/- 3 bpm.

o

o

o

o

4.

Set the simulator to 80%

SpO

2

and 60 PPM.

Verify that after a brief period of
initialization, the

SpO

2

numeric

displays an

SpO

2

value between

77% and 83%.

Verify that the displayed pulse
rate reads 60 +/- 3 bpm.

o

o

o

o

5.

Set the simulator to 90%

SpO

2

and 60 PPM.

Verify that after a brief period of
initialization, the

SpO

2

numeric

displays an

SpO

2

value between

87% and 93%.

Verify that the displayed pulse
rate reads 60 +/- 3 bpm.

o

o

o

o