C

HAPTER

1

M

AINTENANCE

T

ESTS

12

9650-0804-01 Rev. A

Corrective Action

If the unit fails the test, the most likely cause is damaged or disconnected internal tubing. The
device needs to be disassembled to replace or reattach the appropriate segment of NIBP tubing.
The NIBP valves should also be inspected.