Functional Verification Tests
9650-0804-01 Rev. A
ZOLL Propaq M Service Manual
25
Heart Rate Alarm Test
Tools Needed
ECG simulator
Do this...
Observe this...
Pass/Fail
1.
Press the power button on the
Propaq M to turn the unit on.
Connect the ECG leads to the
ECG
simulator
. Set the simulator to 120
BPM and press the Leads quick
access key
(
)
on the Propaq M
unit and select lead II.
II displays next to the Leads quick
access key.
o
o
2.
On the Propaq M:
•
Press the Next quick
access key (
) key to
access the second set of
quick access keys.
•
Press the Alarm quick
access key (
).
•
Press the Limits softkey.
The Alarm Settings menu displays.
o
o
3.
On the Propaq M, select HR/PR
Alarm in the Alarm Settings menu
and use the navigation keys to do
the following:
•
Set the Upper Limit Enable
to ON.
•
Set the Lower Limit
Enable to ON.
•
Set the Lower limit to 30.
•
Set the Upper limit to 150.
Use the navigation keys to exit the
Alarm Settings menu.
4.
On the Propaq M, press the Alarm
Suspend key (
).
The red alarm suspension timer
displays at the top of the next to the
message area.
5.
Remove a lead wire from the
ECG
simulator
.
T
he ECG LEAD FAULT message
displays.
o
o
6.
On the Propaq M, press the Alarm
Suspend key (
).
The Alarm Suspend key flashes
.
7.
Reattach ECG Lead wire to
ECG
simulator
.
The ECG LEAD FAULT message no
longer displays.
o
o
Set simulator to 20 BPM or lower.
Heart Rate Value is highlighted and
the alarm tone sounds. The HR/PR
LOW ALARM message displays.
8.
On the Propaq M, press the Alarm
Suspend key (
).
The Alarm Suspend key flashes.
The
red alarm suspension timer displays
at the top of the screen next to the
message area.
o
o
9.
Set the simulator to 120 BPM.
I, II,
III...