ZOLL Propaq M Rev A User Manual
Page 40
C
HAPTER
2
T
ROUBLESHOOTING
36
9650-0804-01 Rev. A
NIBP MONITORING
FAILURE
One of the following issues
occurred:
•
A failure was reported by
the NIBP module.
•
The two transducers are
not aligned. They are
either not calibrated,
there is a transducer
failure, or there is a
kinked tube.
•
A calibration problem was
reported by the NIBP
module.
If problem persists, replace
the NIBP module.
NIBP COMMUNICATION
FAILURE
There is a communication
problem with the NIBP
module.
Ensure that the J32
connection is secure. If
error persists, replace the
NIBP module.
NIBP CALIBRATION
CHECK DUE
Annual calibration check is
due.
Perform the NIBP
Calibration Check
procedure in Chapter 1,
"Calibration and Calibration
Checks".
NIBP KINKED HOSE
The cuff inflation was too
rapid. There may be kinked
tubing or another air
obstruction.
Check the external NIBP
hose for kinks. If problem
persists, check the routing
of the NIBP tubing inside
the device.
NIBP MAX PRESSURE
EXCEEDED
Pressure limit exceeded.
If problem persists, do the
following:
•
Check the calibration via
the NIBP Calibration
Check procedure in
Chapter 1, "Calibration
and Calibration
Checks".
•
Check the routing of the
NIBP tubing inside the
device for kinks or
obstructions.
If the NIBP module is out of
calibration, replace the
NIBP module.
Error Message
Explanation
Technical Action