QUICK ACCESS SW 1
SHORT
Quick access key 1 is stuck.
Replace the front panel.
QUICK ACCESS SW 2
SHORT
Quick access key 2 is stuck.
Replace the front panel.
QUICK ACCESS SW 3
SHORT
Quick access key 3 is stuck.
Replace the front panel.
QUICK ACCESS SW 4
SHORT
Quick access key 4 is stuck.
Replace the front panel.
QUICK ACCESS SW 5
SHORT
Quick access key 5 is stuck.
Replace the front panel.
QUICK ACCESS SW 6
SHORT
Quick access key 6 is stuck.
Replace the front panel.
QUICK ACCESS SW 7
SHORT
Quick access key 7 is stuck.
Replace the front panel.
SCREEN CYCLE SHORT
Display/home key is stuck.
Replace the front panel.
ALARM SILENCE SHORT
Alarm Silence/Reset key is
stuck.
Replace the front panel.
NAV PAD UP SHORT
Up/Clockwise Navigation
key is stuck.
Replace the front panel.
NAV PAD SELECT SHORT
Select key is stuck.
Replace the front panel.
NAV PAD DOWN SHORT
Down/Counterclockwise
Navigation key is stuck.
Replace the front panel.
SNAPSHOT SHORT
Snapshot key is stuck.
Replace the front panel.
ECG DEVICE FAILURE
ECG software detected a
problem with the ECG data
that is most likely a hardware
issue.
Replace the main board.
TEMP DEVICE FAILURE
Temp software detected a
problem with the
temperature data that is
most likely a hardware issue.
Replace the main board.
TEMP PROBE1 SHORTED
An A/D value was detected
that indicated the probe was
shorted.
Replace the temperature
probe or adapter cable. If
problem persists, replace
the right side panel.
Error Message
Explanation
Technical Action