ZOLL Propaq M Rev A User Manual
Page 42
C
HAPTER
2
T
ROUBLESHOOTING
38
9650-0804-01 Rev. A
NIBP NEONATAL
DETECTED
Device thinks a neonate cuff
is connected. If no neonate
cuff is connected, tubing
may be damaged or kinked.
Verify that neonate cuff is
not connected. If problem
persists, check the routing
of the NIBP tubing inside
the device.
NIBP LOW VOLTAGE
Insufficient voltage is
supplied to the NIBP
module.
Check the module voltage
in the NIBP Test screen.
Inspect the unit or water
ingress. If evidence of
water ingress is found,
allow the water to
evaporate while periodically
checking the module
voltage. If problem persists,
replace the NIBP module.
SPO2 COM ERROR CODE
Communication error.
Check communication
cables. If problem persists,
replace the SpO
2
module.
SPO2 SENSOR FAILURE
Defective sensor or sensor
error.
Replace the SpO
2
sensor.
SPO2 UNKNOWN
RECOVERY NUM
Recovery error.
Replace the SpO
2
sensor.
CO2 DEVICE FAILURE
Malfunction or device failure.
Replace the CO
2
module.
CO2 COM ERROR CODE
Communication failure.
Check communication
cables. If problem persists,
replace the CO
2
module.
CO2 BAROMETRIC OUT
OF RANGE
Altitude is too high for
module.
Bring device into specified
altitude range. If problem
persists, replace the CO
2
module.
CO2 CAL REMINDER
Calibration reminder.
Calibrate the CO
2
module
soon.
CO2 CAL REQUIRED
Calibration due.
Calibrate the CO
2
module.
CO2 CHECK FLOW
Exhaust port blocked.
Check for exhaust port
occlusion. If problem
persists, replace the CO
2
module.
Error Message
Explanation
Technical Action