38

9650-0804-01 Rev. A

NIBP NEONATAL
DETECTED

Device thinks a neonate cuff
is connected. If no neonate
cuff is connected, tubing
may be damaged or kinked.

Verify that neonate cuff is
not connected. If problem
persists, check the routing
of the NIBP tubing inside
the device.

NIBP LOW VOLTAGE

Insufficient voltage is
supplied to the NIBP
module.

Check the module voltage
in the NIBP Test screen.
Inspect the unit or water
ingress. If evidence of
water ingress is found,
allow the water to
evaporate while periodically
checking the module
voltage. If problem persists,
replace the NIBP module.

SPO2 COM ERROR CODE

Communication error.

Check communication
cables. If problem persists,
replace the SpO

2

module.

SPO2 SENSOR FAILURE

Defective sensor or sensor
error.

Replace the SpO

2

sensor.

SPO2 UNKNOWN
RECOVERY NUM

Recovery error.

Replace the SpO

2

sensor.

CO2 DEVICE FAILURE

Malfunction or device failure.

Replace the CO

2

module.

CO2 COM ERROR CODE

Communication failure.

Check communication
cables. If problem persists,
replace the CO

2

module.

CO2 BAROMETRIC OUT
OF RANGE

Altitude is too high for
module.

Bring device into specified
altitude range. If problem
persists, replace the CO

2

module.

CO2 CAL REMINDER

Calibration reminder.

Calibrate the CO

2

module

soon.

CO2 CAL REQUIRED

Calibration due.

Calibrate the CO

2

module.

CO2 CHECK FLOW

Exhaust port blocked.

Check for exhaust port
occlusion. If problem
persists, replace the CO

2

module.

Error Message

Explanation

Technical Action