EZCT S2 USER’S MANUAL

REV 3

52

5.3

Verifying the EZCT S2’s I

x

Sense Circuit Using an External Meter

You can verify the excitation current (I

x

) sensed by the EZCT S2 by using an external RMS

ampere meter. Follow the steps below to verify the EZCT S2’s I

x

sense circuit:

a. Connect the X cables to a power resistor and an RMS ampere meter as shown in Figure

14.

b. Select the EZCT S2’s Diagnostics mode (see section 5.1).

c. Press the

[PAPER Contrast]

key to raise the V

x

voltage.

d. Verify the I

x

voltage displayed on the EZCT S2’s LCD screen against the value displayed

on the external RMS ampere meter.

e. Press the

[STOP]

key to end the test.

Figure 14. EZCT S2 I

x

Verification Test Connections