beautypg.com

Ct-7000 operating instructions – Vanguard CT-7000 User Manual

Page 17

background image

CT-7000 Operating Instructions

10

7.2 CT-7000 Operating Voltages

The CT-7000 operating voltage is selectable between 110/120 Vac, 50/60 Hz or 220/240

Vac, 50/60 Hz. Voltage selection is set by the JP1 connector on the Power Supply board and the
JP3 connector on the Initiate board of the CT-7000. The jumper settings for the connectors are
shown in Table 3.0.

Table 3.0 Voltage Selection Jumper Setting

VOLTAGE SELECTION

CONNECTOR PIN

110-120 Vac

Pin 1& 2, 3 & 4

220-240 Vac

Pin 2 & 3

An initiate circuit relay with the corresponding operating voltage must also be installed in the

CT-7000 Initiate board.

NOTE

The factory sets the operating voltage.

7.3 CT-7000 Main Power Fuse

The CT-7000 uses an AC input module that contains the AC receptacle, power switch and

protective fuses. The 3 Ampere replacement fuses should be 20mm, 250Vac, fast-blow types.

NOTE

The OPEN and CLOSE initiate are 5 Ampere fuses that should be 3AG, 250Vac,
fast-blow types.

7.4 CT-7000 Printer and Printer Paper

The built-in, thermal printer uses 4.5 inch wide, thermal paper for printing test results. To

maintain the highest, quality printing and to avoid paper jamming, it is highly recommended that
the thermal paper be supplied by the factory. Additional paper can be ordered from either of the
two sources listed below:

Vanguard Instruments Co, Inc.
1710 Grevillea Court
Ontario, CA 91761
Tel: 909-923-9390
Fax: 909-923-9391
Part Number: TP-4 Paper

BG Instrument Co.
13607 E. Trent Avenue
Spokane, WA 99216
Tel: 509-893-9881
Fax: 509-893-9803
Part Number: TP-4 paper