QUERY SYNTAX

:MoDulateWaVe?

:=

{C1|C2}

RESPONSE

:MoDulateWaVe ,,…

:= {any parameters set for the modulated waveform}

EXAMPLE 1

Set Channel 1 modulation to AM, and AM wave shape to SINE wave.

C1:MDWV STATE,ON,AM,MDSP,SINE

EXAMPLE 2

Set Channel 1 FM frequency to 1000Hz.

C1:MDWV STATE,ON,FM,FRQ,1000HZ

EXAMPLE 3

Read Channel 1 Modulate Waveform parameters (and STATE is ON).

C1:MDWV?

Return:

C1:MDWV STATE,ON,AM,MDSP,SINE,SRC,INT,FRQ,100HZ,

DEPTH,100,CARR,WVTP,RAMP,FRQ,1000HZ,AMP,4V,OFST,0V,SYM,50

RELATED

ARWV, BTWV, CFG, CPL, SWWV, BSWV

22