Parameters

Value

Description

(CARR,)AMP

Where carrier wave := {default unit is V;

WS2000 Channel 1 min. 0.004V max. 6V, Channel 2 min. 0.004V max. 20V

WS3000 Channel 1 and 2 min. 2mVpp max. 20Vpp}.

(CARR,)OFST

Where carrier wave := {default unit is V; maximum value depends on

the WaveStation model}.

(CARR,)SYM

Where carrier wave := {0% to 100%}.

(CARR,)DUTY

Where carrier wave := {if CARR,WVTP is SQUARE, 20% to

80%; if CARR,WVTP is PULSE, 0.1% to 99.9%}.

(CARR,)PHSE

Where carrier wave := {0 to 360}.

(CARR,)RISE

Value of carrier wave rising edge. Parameter valid only when CARR,WVTP

is PULSE.

(CARR,)FALL

Value of carrier wave falling edge. Parameter valid only when CARR,WVTP

is PULSE.

(CARR,)DLY

Where carrier wave := {default unit is S; minimum 0, maximum is

value of Pulse Period}. Parameter is valid only when CARR,WVTP is

PULSE.

(CARR,)STDEV

deviation>

Parameter is valid only when CARR,WVTP is NOISE.

(CARR,)MEAN

Parameter is valid only when CARR,WVTP is NOISE.

QUERY SYNTAX

:BursTWaVe?

RESPONSE

:BursTWaVe ,,

EXAMPLE 1

Set Channel 1 burst waveform period to one second.

C1:BTWV PRD,1S

EXAMPLE 2

Read Channel 2 burst waveform parameters (and STATE is ON).

C2:BTWV?

Return:

C2:BURSTWAVE STATE,ON,PRD,0.01S,STPS,0,TRIG,IN,TRMD,OFF,

TIME,1,DLAY,2.4e-07S,GATE_NCYC,NCYC,PLRT,NEG,CARR,WVTP,SINE,

FRQ,1000HZ,AMP,4V,OFST,0V,PHSE,0

EXAMPLE 3

Read Channel 2 burst wave parameters (and STATE is OFF).

C2:BTWV?

Return:

C2:BURSTWAVE STATE,OFF

18