Teledyne LeCroy WaveStation SCPI Command Reference Manual User Manual
Page 20
![background image](/manuals/353697/20/background.png)
WaveStation SCPI
Parameters
Value
Description
(CARR,)AMP
Where carrier wave
WS2000 Channel 1 min. 0.004V max. 6V, Channel 2 min. 0.004V max. 20V
WS3000 Channel 1 and 2 min. 2mVpp max. 20Vpp}.
(CARR,)OFST
Where carrier wave
the WaveStation model}.
(CARR,)SYM
Where carrier wave
(CARR,)DUTY
Where carrier wave
80%; if CARR,WVTP is PULSE, 0.1% to 99.9%}.
(CARR,)PHSE
Where carrier wave
(CARR,)RISE
Value of carrier wave rising edge. Parameter valid only when CARR,WVTP
is PULSE.
(CARR,)FALL
Value of carrier wave falling edge. Parameter valid only when CARR,WVTP
is PULSE.
(CARR,)DLY
Where carrier wave
value of Pulse Period}. Parameter is valid only when CARR,WVTP is
PULSE.
(CARR,)STDEV
deviation> Parameter is valid only when CARR,WVTP is NOISE. (CARR,)MEAN Parameter is valid only when CARR,WVTP is NOISE. QUERY SYNTAX RESPONSE EXAMPLE 1 Set Channel 1 burst waveform period to one second. C1:BTWV PRD,1S EXAMPLE 2 Read Channel 2 burst waveform parameters (and STATE is ON). C2:BTWV? Return: C2:BURSTWAVE STATE,ON,PRD,0.01S,STPS,0,TRIG,IN,TRMD,OFF, TIME,1,DLAY,2.4e-07S,GATE_NCYC,NCYC,PLRT,NEG,CARR,WVTP,SINE, FRQ,1000HZ,AMP,4V,OFST,0V,PHSE,0 EXAMPLE 3 Read Channel 2 burst wave parameters (and STATE is OFF). C2:BTWV? Return: C2:BURSTWAVE STATE,OFF 18 922829 Rev C