3 example 3: cascading two sierra m6-4 systems, Igure, Xample – Teledyne LeCroy Sierra Port Configurations Reference Manual User Manual
Page 12: Ascading, Ierra, M6-4, Ystems

Teledyne LeCroy Corporation Sierra M6-4 and Sierra M6-2 Port Configurations Reference Manual
6
6
1.4.3
Example 3: Cascading Two Sierra M6-4 Systems
Ports configuration:
System 1
Port 1
Port 2
Port 3
Port 4
-
-
-
System 2
Port 1
Port 2
Port 3
Port 4
-
-
-
Setup:
For System 1, Port 1 of Initiator connects to the Host DUT by a mini-SAS cable. Using Jammer allows
error injection on Port 1. Using Analyzer allows analysis on Port 1 before and after error injection.
For System 1, Port 1 of Target connects to Port 2 of the Initiator port of System 2 by a mini-SAS cable.
For System 2, Port 2 of Initiator connects to Port 1 of the Target port of System 1 by a mini-SAS cable.
For System 2, Port 1 of Target connects to a Device Under Test (DUT 1) by a single octopus cable.
For System 2, Port 2 of Target connects to a Device Under Test (DUT 2) by a single octopus cable.
Using Jammer allows error injection on Port 2. Using Analyzer allows analysis on Port 2 before and after
error injection.
Description:
Port 1 of System 1 is configured to jam the incoming traffic and output it back. In addition, the system
is internally connected so that one analyzer on Port 1 is capturing the incoming (and pre-jammed)
traffic, while another analyzer on Port 1 can capture the outgoing (post-jammed) traffic of physical Port
1.
Port 2 of System 2 is configured to jam the incoming traffic and output it back. In addition, the system
is internally connected so one analyzer on Port 2 is capturing the incoming (and pre-jammed) traffic,
while another analyzer on Port 2 can capture the outgoing (post-jammed) traffic of physical Port 2.
This configuration allows impairing traffic on both physical ports of DUT1 and DUT2, while capturing
pre-jammed and post-jammed traffic on each.
Ports 3 and 4 on both systems are not used.
Figure 1-4 Example 3: Cascading Two Sierra M6-4 Systems
DUT 2
DUT 1
Host DUT
Mini-SAS
System 1
System 2
Mini-SAS
cable
Port 2
“Octopus”
cable
Port 1