beautypg.com

Performance profi le – SKC 210-1000 Pocket Pump User Manual

Page 5

background image

3

Performance Profi le

Timer:

1 to 9999 min display ± 1% accuracy

Flow Fault:

If the pump is unable to compensate for longer than
15 seconds due to excessive back pressure, the pump
enters FLOW FAULT mode. The pump goes into
HOLD, the fault icon appears on the display, and the
accumulated run time display is frozen and retained.
Aft er 5 minutes in fl ow fault, auto-restart is att empted
every 5 minutes until fl ow is corrected or the user
removes the pump from fl ow fault mode by pressing
[].

Batt ery Pack:

Rechargeable nickel-metal hydride (NiMH)

batt ery pack, 2.4 V x 1.0 Ah


Rechargeable nickel cadmium (NiCad) batt ery

pack, 2.4 V x 0.6 Ah

Charge Time:

Att ached to pump: ≤ 6 hrs

Detached from pump: 16 hrs (recommended charging
method for initial charge of a new batt ery pack
)

Size:

4.5 x 2.2 x 1.4 in (11.4 x 5.6 x 3.6 cm) - see photo below

Weight:

5 oz (142 gm)

The use of a repaired or rebuilt battery pack voids any warranty and
the UL Listing for intrinsic safety.

2.2 inches

4.5 inches

1.4 inches

Newer pump models contain an NiMH
battery. An NiMH battery can be used with
older Pocket Pump models. Order NiMH
Battery Pack Cat. No. P20129-2.

Pocket Pump was designed to minimize
inherent noise. If the noise level is
distracting when worn in a shirt pocket,
clip the pump on a waist belt.

Note:

Note:

(varies with batt ery capacity
and level of discharge)