Operation, Calibrating and sampling with bags – SKC 210-1000 Pocket Pump User Manual

Page 21

background image

19

Operation

Calibrating and Sampling with Bags

Requires Twin Port Pocket Pump Cat. No. 210-1002A

Calibration

Before use, allow the pump to equilibrate after moving it from one temperature
extreme to another.

1. Ensure the pump is set in Constant Flow mode (see pages 12 and 13).

2. Use tubing to connect a primary standard calibrator to the pump inlet.

3. Press

[] to RUN the pump.

4. Press



(security code) within 10 seconds. The word SET will fl ash on

the LCD.

5. The LCD will display the fl ow rate set from the last sample taken. If you do not

wish to change the fl ow rate, go to Step 7.

6. Press

 or  to change the fl ow rate to the desired setting as shown on the

LCD.

7. Press

. ADJ will appear on the LCD. The pump fl ow rate can now be

calibrated using a calibrator. See reminder below.

8. When

the

fl ow rate reading appears on the calibrator, press the  or  buttons

on the pump keypad to adjust the fl ow up or down until the calibrator displays
the desired fl ow rate.

Twin port Pocket Pump

sampling train with bag

Outlet
(exhaust)

Inlet

When calibrating, the fl ow rate displayed on the calibrator will change as a
result of this adjustment, not the fl ow rate displayed on the pump.