Operation, Calibrating and sampling with bags – SKC 210-1000 Pocket Pump User Manual
Page 21
19
Operation
Calibrating and Sampling with Bags
Requires Twin Port Pocket Pump Cat. No. 210-1002A
Calibration
Before use, allow the pump to equilibrate after moving it from one temperature
extreme to another.
1. Ensure the pump is set in Constant Flow mode (see pages 12 and 13).
2. Use tubing to connect a primary standard calibrator to the pump inlet.
3. Press
[] to RUN the pump.
4. Press
(security code) within 10 seconds. The word SET will fl ash on
the LCD.
5. The LCD will display the fl ow rate set from the last sample taken. If you do not
wish to change the fl ow rate, go to Step 7.
6. Press
or to change the fl ow rate to the desired setting as shown on the
LCD.
7. Press
. ADJ will appear on the LCD. The pump fl ow rate can now be
calibrated using a calibrator. See reminder below.
8. When
the
fl ow rate reading appears on the calibrator, press the or buttons
on the pump keypad to adjust the fl ow up or down until the calibrator displays
the desired fl ow rate.
Twin port Pocket Pump
sampling train with bag
Outlet
(exhaust)
Inlet
When calibrating, the fl ow rate displayed on the calibrator will change as a
result of this adjustment, not the fl ow rate displayed on the pump.