Setup, Setting up the pump – SKC 210-1000 Pocket Pump User Manual
Page 14
12
Setup
Pump run time is a converse relationship: Low fl ow + low Bp = longer run time
High
fl ow + high Bp = shorter run time
Note:
Setting Up the Pump
To activate the pump:
• Press any keypad button. The LCD will show the pump serial number for two
seconds followed by a fi rmware revision number.
To obtain run time data:
•
Press
the
button repeatedly to scroll through run time data.
To reset the data display to zero:
• With the pump running, press:
[] to place in HOLD, to enter Setup, and .
The LCD will show briefl y the pump serial number, fi rmware version number,
and run time at zero minutes. This will reset all run time data except the fl ow
setting. See instructions on page 15 to change the fl ow setting. If you do not
want to reset the run time data after entering the security code, stop and wait
10 seconds to break the sequence. The LCD will stop fl ashing at this point.
To determine pump operating mode (constant fl ow or constant
pressure):
• Observe the LCD. Use the button to scroll through the displays if necessary.
Constant fl ow mode
Constant pressure mode (Multiple tube)
1 to 3 digit number is displayed
with “mL/min” and “FLOW”
1 to 3 digit number is displayed
with “P” and “ins” or “mm”