Operation – SKC 210-1000 Pocket Pump User Manual
Page 18
16
Operation
9. Press
. ADJ will appear on the LCD. The pump fl ow rate can now be calibrated
using a calibrator. See reminder below.
10. When the fl ow rate reading appears on the calibrator, press the or buttons
on the pump keypad to adjust the fl ow up or down until the calibrator displays
the desired fl ow rate.
11. Press to lock in the calibrated fl ow. The pump will then enter its normal RUN
state.
12. Press [] to place the pump in HOLD. Press
(security code) and
to reset the data display to zero (minutes that have elapsed during pump setup
and/or programming) before sampling.
13. Disconnect the calibrator and replace the representative sample tube with the tube
to be used for sampling.
Sampling
1. Clip the sampling medium to a worker’s collar
in the breathing zone and place the pump in the
worker’s shirt pocket or clip to the worker’s
belt.
2. Press
[] to RUN the pump and begin
sampling.
3. Press [] to HOLD the pump and stop
sampling. Run time data is retained in memory
after sampling is completed. While the pump
is in HOLD mode, use the button to scroll
through run time data on the LCD.
4. Remove and cap the tube. Replace the representative tube in the holder and
reconnect the calibrator. Verify that the fl ow rate has remained within 5% of the
pre-sample calibrated fl ow.
When calibrating, the fl ow rate displayed on the calibrator will change as a
result of this adjustment, not the fl ow rate displayed on the pump.
Protect the sample pump from weather when in use outdoors.
Use of any device other than the approved battery pack to power the pump
voids the UL Listing for intrinsic safety.