beautypg.com

Operation – SKC 210-1000 Pocket Pump User Manual

Page 20

background image

18

Operation

5. Label all tubes and ports (e.g., tube #1, port A).

6. Insert opened representative tubes into the ports. Place unopened tubes in any

unused ports to “seal” them. This is essential to obtain correct results.

7. Loosen

the

fl ow adjust screw on the low fl ow tube holder port containing the

tube to be calibrated. Connect the exposed end of the tube to a primary standard
calibrator using another length of tubing.

8. Press

[] to RUN the pump.


9. Turn

the

fl ow adjust screw (needle valve) on the adjustable tube holder port until

the calibrator displays the desired fl ow rate through the tube. The fl ow rate dis-
played on the calibrator changes as a result of this adjustment.

10. To calibrate fl ow through the remaining tubes, repeat Steps 7 through 9 for each

port.

11. Press [] to place the pump in HOLD mode. Press  (security code)

and  to reset the data display to zero before sampling.

12. Disconnect the calibrator and replace the representative tubes with the tubes to

be used for sampling. The pump is ready to sample.

Sampling

1. Clip the sampling medium to a worker’s collar in the breathing zone and place

the pump in the worker’s shirt pocket or clip to the worker’s belt.

2. Press

[] to RUN the pump and begin sampling.

3. Press

[] to HOLD the pump and stop sampling.

Run time data is retained in memory after sampling is completed. While the pump is
in HOLD mode, use the  button to scroll through run time data on the LCD.

4. Remove and cap tubes. Reconnect calibration train and verify the fl ow rate.

All empty ports must contain unopened tubes during calibration and sampling.

Protect the sample pump from weather when in use outdoors.

Do not charge or operate the pump with the charger in hazardous locations.

Use of any device other than the approved battery pack to power the pump
voids the UL Listing for intrinsic safety.