Performance profi le – SKC 210-1000 Pocket Pump User Manual
Page 4
2
Performance Profi le
• RFI/EMI shielded
• CE marked
•
Listed for intrinsic safety (single port models only)
• Model 210-1002TX is ATEX approved
Flow Range in Constant
Flow Mode:
20 to 225 ml/min
Accuracy Variance Between
LCD Reading and Actual Flow
Rate (aft er calibration):
20 to 225 ml/min ± 5%
Constant Flow Compensation
Range (inlet only):
20 to 225 ml/min up to 20 inches water back pressure
Flow Control:
Holds constant fl ow to ± 5% of the set-point
Pressure Range in Constant
• 1 to 10 inches water (1.87 to 18.7 mm Hg) at
Pressure (multiple-tube
maximum
fl ow rate of 200 ml/min
sampling) Mode (inlet only):
• 10 to 20 inches water (18.7 to 37.4 mm Hg) at
maximum
fl ow rate of 100 ml/min
Constant Pressure (multiple-
tube sampling) Mode Accuracy: Pressure reading ± 0.5 inches water (0.25 mm Hg)
Batt ery Charge Level Indicator: Icon displays at full, mid, and low charge
Temperature Range:
Operating: 32 to 113 F (0 to 45 C)
Charging: 40 to 100 F (5 to 38 C)
Storage: -4 to 113 F (-20 to 45 C)
Operating Humidity:
0 to 95% non-condensing
Protect sample pump from weather when in use outdoors.
Tubing:
Requires 1/4-in ID tubing
Run Time:
• With NiMH batt ery: 12 hrs at 200 ml/min up to
(dependent on sample media used) 10 inches water back pressure
12 hrs at 100 ml/min up to 20 inches water back
pressure
(Run time is 8 hrs at either sett ing if using a
NiCad
batt ery.)
• Connected to charger: Extended run time