Performance Profi le

• RFI/EMI shielded
• CE marked

Listed for intrinsic safety (single port models only)

• Model 210-1002TX is ATEX approved

Flow Range in Constant
Flow Mode:

20 to 225 ml/min

Accuracy Variance Between
LCD Reading and Actual Flow
Rate (aft er calibration):

20 to 225 ml/min ± 5%

Constant Flow Compensation
Range (inlet only
):

20 to 225 ml/min up to 20 inches water back pressure

Flow Control:

Holds constant fl ow to ± 5% of the set-point

Pressure Range in Constant

1 to 10 inches water (1.87 to 18.7 mm Hg) at

Pressure (multiple-tube

maximum

fl ow rate of 200 ml/min

sampling) Mode (inlet only):

10 to 20 inches water (18.7 to 37.4 mm Hg) at

maximum

fl ow rate of 100 ml/min

Constant Pressure (multiple-
tube sampling) Mode Accuracy:
Pressure reading ± 0.5 inches water (0.25 mm Hg)

Batt ery Charge Level Indicator: Icon displays at full, mid, and low charge

Temperature Range:

Operating: 32 to 113 F (0 to 45 C)

Charging: 40 to 100 F (5 to 38 C)

Storage: -4 to 113 F (-20 to 45 C)

Operating Humidity:

0 to 95% non-condensing

Protect sample pump from weather when in use outdoors.

Tubing:

Requires 1/4-in ID tubing

Run Time:

• With NiMH batt ery: 12 hrs at 200 ml/min up to

(dependent on sample media used) 10 inches water back pressure

12 hrs at 100 ml/min up to 20 inches water back
pressure

(Run time is 8 hrs at either sett ing if using a

NiCad

batt ery.)

• Connected to charger: Extended run time