Operation – SKC 210-1000 Pocket Pump User Manual

Page 22

background image

20

Operation

9. Press

to lock in the calibrated fl ow. The pump will then enter its normal

RUN state.

10. Press [] to place the pump in HOLD. Press  (security code) and



to reset the data display to zero (minutes that have elapsed during pump

setup and/or programming) before sampling.

11. Disconnect the calibrator and tubing.

Sampling

1. Ensure the data display has been reset to zero (see page 12).

2. Using a length of PTFE tubing, connect the pump outlet (exhaust) to the hose

connection on the fi tting of a prepared sample bag. Ensure shut-off valve is open
on the bag fi tting. See sample bag operating instructions for details.

3. Press

[] to RUN the pump. Sample for the appropriate length of time.

Do not over-infl ate the sample bag. Avoid fi lling the bag more than 80% of its
maximum volume.

4. When sampling is complete, press [] to place the pump in HOLD and stop

sampling. Close the shut-off valve on the sample bag fi tting. See sample bag
operating instructions for details.

Incorrect

Infl ation

Correct

Infl ation