Operation
9. Press
to lock in the calibrated fl ow. The pump will then enter its normal
RUN state.
10. Press [] to place the pump in HOLD. Press (security code) and
to reset the data display to zero (minutes that have elapsed during pump
setup and/or programming) before sampling.
11. Disconnect the calibrator and tubing.
Sampling
1. Ensure the data display has been reset to zero (see page 12).
2. Using a length of PTFE tubing, connect the pump outlet (exhaust) to the hose
connection on the fi tting of a prepared sample bag. Ensure shut-off valve is open
on the bag fi tting. See sample bag operating instructions for details.
3. Press
[] to RUN the pump. Sample for the appropriate length of time.
Do not over-infl ate the sample bag. Avoid fi lling the bag more than 80% of its
maximum volume.
4. When sampling is complete, press [] to place the pump in HOLD and stop
sampling. Close the shut-off valve on the sample bag fi tting. See sample bag
operating instructions for details.
Incorrect
Infl ation
Correct
Infl ation