Calibrating and Sampling in Constant Flow Mode —
Single Sorbent Tube

Calibration

Before use, allow the pump to equilibrate after moving it from one temperature
extreme to another.

1. Ensure the pump has run for fi ve minutes before proceeding with calibration.

2. Set the pump for Constant Flow mode (see pages 12 and 13).

3. Using a length of tubing, connect the pump inlet to a representative sorbent tube

using an SKC tube holder.

4. Connect the exposed end of the tube to a primary standard calibrator using another

length of tubing.

5. Press [] to RUN the pump.

6. Press  (security code) within 10 seconds. The word SET will fl ash on the

LCD.

7. The LCD will display the fl ow rate set

from the last sample taken. If you do
not wish to change the fl ow rate, go to
Step 9.

8. Press  or  to change the fl ow rate to

the desired setting as shown on the LCD.

Do not charge or operate the pump with the charger in hazardous locations.

Operation