Setup – SKC 210-1000 Pocket Pump User Manual

Page 15

background image

13

Setup

To change pump operating mode (constant fl ow or constant

pressure):

• With the pump running, press:

[] to place in HOLD,  to enter Setup, and



The display will now show the new operating mode.

For information on operating the pump in constant pressure mode,
see Advanced Operation on page 21.

To determine the pump display:

Standard display includes fl ow, volume, and run time. Press  to scroll

through run time data.

Enhanced Display includes fl ow, volume, back pressure, temperature, and

run time. Press  to scroll through run time data.

To select the pump display:

Standard to Enhanced Display:

With the pump running, press:

[] to place in HOLD,  to enter Setup, and



The enhanced data display will include fl ow, volume, back pressure, temperature,

and

run time.

Enhanced to Standard Display:

With pump running, press:

[] to place in HOLD,  to enter Setup, and



The standard data display will include fl ow, volume, and run time.

To select the temperature scale display:

• In enhanced display, the temperature of incoming air can be shown in either

Celsius (C) or Fahrenheit (F). Use the  button to scroll to the temperature
scale display units.