Robot waterbase applicator - maintenenace, Fluid system – Ransburg Evolver Water Applicator 78863-90, 78863-60 User Manual
Page 39

35
Robot Waterbase Applicator - Maintenenace
General Problem
Possible Causes
Corrective Action
Excessive Current or Loss
of High Voltage
1. Fluid tube leak inside fluid
isolation tube, charged paint
grounds out at robot arm
manifold.
2. Tubing or conductive
sheath nicked or cut.
3. Internal end of fluid tube not
seated properly into valve
manifold assembly.
4. Contaminated dump line.
5. Exterior of gun contaminat-
1. Replace any worn or damaged o-rings on
EMFD-29-1 barbed fittings.
2. Replace any worn or damaged o-rings on fluid
isolation tube.
3. Fluid tube too short. Check for correct length
of fluid tube. Refer to Figure 21, Robot
Manifold Assembly.
1. Replace tubing or conductive sheath.
1. Ensure fluid tube is seated tightly into valve
manifold assembly.
1. Clean or replace
2. Clean with non-polar solvent
Fluid System
Fluid and Isolation Tube
Note: To check for fluid leaks, it is easiest to first remove the gun from the test station and remove the shroud
from the gun. The aluminum flange can be reattached to the gun manifold plate by using the three stainless
steel screws. Then the gun can be remounted to the test station. Leak detector may be used at all appropriate
sources. Be certain to wipe off all residual solution using a non-conductive solvent such as Naphtha.
AA-03-01.1