Dissolved Oxygen Controller

Operation

Teledyne Analytical Instruments

49

2 DO

10.0 ppm

Cal

-157.5 mV/FS


Standardization menu @ factory default

settings

1 DO

.00 ppm

Cal

.0 mV


Standardization menu DO @ factory default

settings

4. Immediately before making the calibration adjustment, note the

current value on the display and add the differential to the
current value. This "total adjusted" value is entered as the buffer
value.

3.4.2.5

C

ALIBRATION

E

RROR

D

ETECTION

& E

LECTRODE

O

PERATIONAL

G

UIDELINES

If the slope calibration performed in Section 3.4.2.3 is not within

the predetermined limits of the instrument, a flashing asterisk (*) will
appear.

The presence of the asterisk indicates a potential calibration

problem. In general, if the asterisk appears, the integrity of the
electrode, the buffer solutions or the handling procedure should be
questioned. See the dissolved oxygen sensor manual for specifics on the
electrode.

3.4.2.6 B

ACK

-

TO

-F

ACTORY

C

ALIBRATION

This feature allows the LXT-220 to be reset to "typical electrode"

for electrode potentials. As the electrode ages it will deteriorate and the
mV/FS will decay and will eventually require replacement. If the
membrane is torn or coated the calibration will be either sluggish or out
of range.

The following procedure for returning the LXT-220 to factory

calibration can be used for both standardize and span calibrations.

Procedure at the Standardization Menu:

1. Position the cursor under the "C" in "Cal."
2. Press both horizontal CALIBRATE keys simultaneously and

observe the cursor briefly move to the right. The default setting
is complete.