Operation

LXT-220 DO

Teledyne Analytical Instruments

46

Buffer
Setup
Status

Parameter Selection Menu

1DO

.00 ppm

Cal

.0 mV


Standardization menu pH

• BUFFER: Perform a 1 or a 2 point calibration

• SETUP: Set up the range of the 4-20 mA output

• STATUS: Information only. Shows the data stored from the last

calibration.

3.4.2.1 B

UFFER

M

ENU

From the parameter selection menu chose the top line (the

BUFFER command) and press the two horizontal ◄► CALIBRATE
keys simultaneously to enter the BUFFER calibration submenu. Both
the one-point calibration (standardize) and the two-point (span)
calibrations are available within this menu. The procedures for each
calibration function are given below.

3.4.2.2 O

NE

-P

OINT

B

UFFER

C

ALIBRATION

(S

TANDARDIZE

)

Because all dissolved oxygen electrodes experience minor

variations, buffer calibrations are necessary before installing the
electrodes in service. Also, occasional calibrations are necessary to
compensate for electrode degradation while in service. For
measurement specifics refer to the dissolved oxygen sensor manual.

Procedure:

1. Be sure that the cursor is located under the "1" as illustrated.
2. For diagnostic reasons note the ppm and mV values from the

prior calibration.