Dissolved Oxygen Controller

Operation

Teledyne Analytical Instruments

35

The configuration password is rarely used and the configuration

settings are more sensitive, so the password should use more digits than
the operational password. With eight base 4 digits the odds of breaking
the code in three tries are 3 in 65536 if the user knows that there are 8
digits. The user will not know the number of digits that you chose, so
the odds are usually smaller than that. The maximum number of digits
allowed in a password, or the number actually used, cannot be
determined by entering trial passwords. A two or three digit password is
therefore moderately effective if systematic attempts at breaking the
code are not anticipated, but short passwords should nevertheless be
avoided in a critical process.

3.3.2 4-20 mA Output Assignments

The next group of displays are used to assign the 4-20 mA outputs.

Note: The LXT-220 has been configured at the factory for your

specific application as specified at the time of the order.

Using the MENU SELECT ▲/▼ keys, scroll down to the 4-20 mA

ASSIGN option and select it by pressing both CALIBRATE ◄/ ► keys
simultaneously. The screen will display 4-20 1: where . . . is one of the
following selections:

missing

not installed

unused

Installed but not currently used

ch1 PV

Channel 1 process value

ch1 temp

Channel 1 temperature

ch1 PID

Channel 1 PID controller output

ch2 PV

Channel 2 process value

ch2 temp

Channel 2 temperature

ch2 PID

Channel 2 PID controller output


The channel 2 options are only shown on two channel instruments,

and specialized instruments may show additional options.

To change the output assignment, first push both CALIBRATE ◄/

► keys, causing the cursor to move to the right. Then use the right-side
CALIBRATE ▲/▼ keys to scroll through the list of options. The list is