background image

Dissolved Oxygen Controller

Operation

Teledyne Analytical Instruments

45

DO

7.80

M 50.0%

25.0

O

C


Main Menu, DO

Procedure:

1. If not at the Main Menu, press the appropriate MENU

SELECTION key to reach the Main Menu.

2. Press both horizontal ◄►CALIBRATE keys simultaneously to

enter the calibrate mode and observe the "M" appearing in front
of the % current output value. The "M" signifies that the manual
output is locked in the manual mode.

3. Using the horizontal CALIBRATE keys one at a time, position

the cursor under the digit to be adjusted.

4. Increase or decrease the value by pressing the appropriate

vertical ▲/▼ CALIBRATE key.

To release manual output:

1. Return to the Main Menu by pressing the appropriate MENU

SELECTION key.

2. Observe that the "M" appears in front of the % current output

value. This "M" signifies that the transmitter output is in
manual.

3. Press only the left-hand ◄ CALIBRATE key and observe the

"M" disappear, releasing the LXT-220 from the manual output
mode. The % output value will return to the real-time current
output.

3.4.2 Parameter Selection Menu

Scroll down from the main menu to reach the parameter set up

menu. Once in the Parameter set up menu select the function you would
like to enter.