beautypg.com

Teledyne LXT-220 User Manual

Page 48

background image

Operation

LXT-220 DO

Teledyne Analytical Instruments

48

3. For diagnostic reasons note the DO and mV/FS values from the

prior calibration.

4. Press both horizontal CALIBRATE keys simultaneously to enter

the calibrate mode. Observe the diagnostic value on the bottom
line becomes a real-time value and that the term "CAL" has
disappeared.

5. To change the buffer point, position the cursor under the desired

digit using the left-hand CALIBRATE key.

6. Increase or decrease the value by pressing the appropriate

vertical CALIBRATE key.

7. Wait for the real-time mV value to stabilize.
8. To file the calibration and exit the calibrate mode, press either

MENU SELECTION key.

A perfect electrode slope (efficiency) is typically greater than -

150.0 mV/FS. If the value falls below -100.0 mV/FS, the electrode
should be serviced or replaced.

IMPORTANT: If a flashing asterisk appears during or after

calibration, the electrode may require service, or
the buffer solution may be contaminated.

Note: Comparing the millivolt values before and after the

calibration provides a valuable diagnostic tool in
determining the degradation of the electrode. Large
differences between calibrations may indicate coating or
damage to the measurement half-cell.

3.4.2.4 C

ALIBRATION

U

SING A

G

RAB

S

AMPLE AS A

S

TANDARD

The LXT-220 allows easy standardization to a grab sample value

by placing the process value into the first calibration point.

Procedure:

1. Record the O

2

value of the process when the sample is extracted.

2. Record the O

2

value of the grab sample.

3. Calculate the difference between the two values as follows:

(Grab sample value) - (Displayed value) = differential