Operation

LXT-220 DO

Teledyne Analytical Instruments

30

Fri 12:00:37



(volatile)

Time of day clock

Input -150.4 mVa
0 mV .00ppm

Electrode Data menu ORP

Exit

Stay


Exit / Stay menu

3.2.4.4 E

XIT

/S

TAY

M

ENU

This menu confirms that the user wishes to leave the section that

they are currently in. This menu will show up each time that the LX-

220 is prompted to leave the current section before it goes back to the
parameter selection menu. Press the appropriate menu selection key
either to stay in the current section or exit and return to the parameter
selection menu.

3.3 Configuring the LXT-220

To initially configure or reconfigure the instrument, use the left-

side ▲ and ▼ MENU SELECT keys to scroll to the bottom of the
menus list and stop when the CONFIGURE/TRIM option is displayed.
Simultaneously depress the◄ and ► keys to enter the configuration
mode. This mode should be password protected if access to the keyboard
is unrestricted. The new configuration remains in effect through
indefinite power outages.

The configuration mode is a submenu grouping of 9 or more

functions that can be set by the user to tailor the instrument to a
particular application. The following 7 functions appear in groups of 4
for a standard instrument:

PASSWORDS
4-20 ASSIGN
4-20 MANUAL MODE
RELAY ASSIGN/TST
°C/°F & TEMP CAL
4-20 TRIM/TEST