Dissolved Oxygen Controller

Operation

Teledyne Analytical Instruments

41

4mA1 trim

.000



4 to 20 mA trim


The bottom line shows the measured temperature. To change the

displayed temperature so it agrees with a measured value, place the
cursor on the top line then simultaneously depress both CALIBRATE
◄/ ► keys. Use the right-side CALIBRATE ▲/▼ keys to adjust the
number so that the displayed temperature agrees with that measured by a
thermometer.

The number shown on the top line is simply added to the

temperature that the instrument measures. There is no slope correction.

To toggle between degrees C and degrees F, place the cursor on the

bottom line and simultaneously depress both CALIBRATE ◄/ ► keys.
This will change the unit. Press both keys again to toggle back to the
original unit.

All temperature displays on both channels use the selected system

of units. The scale and units code in the HART temperature output is
also controlled by this selection.

On two channel instruments the next display trims the channel 2

temperature.

3.3.7 4-20 Trim

The next six displays are used to trim the 4-20 outputs. The 4-20

outputs are calibrated at the factory and will never need recalibration in
most applications. Periodic recalibration is not recommended.


As in the other menus, the cursor moves to the right when both

CALIBRATE keys are depressed and the output current is set to 4.00
mA, overriding the output assignment. Measure the 4-20 output #1
current with an ammeter, and adjust the number on the top line until the
meter reads 4.00 mA.